7: TSP command reference Series 2600B System SourceMeter® Instrument Refe
7-186 2600BS-901-01 Rev. C / August 2016
scriptVar.save()
This function saves the script to nonvolatile memory or to a USB flash drive.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
scriptVar.save()
scriptVar.save(filename)
The name of variable that references the script
The file name to use when saving the script to a USB flash drive
Details
The scriptVar.save() function saves a script to nonvolatile memory or a USB flash drive. The
root folder of the USB flash drive has the absolute path /usb1/.
If no filename is specified (the filename parameter is an empty string), the script is saved to internal
nonvolatile memory. Only a script with filename defined can be saved to internal nonvolatile
memory. If a filename is given, the script is saved to the USB flash drive.
If no filename is specified (the filename parameter is an empty string), the script is saved to internal
nonvolatile memory. Only a script with filename defined can be saved to internal nonvolatile
memory. If a filename is given, the script is saved to the USB flash drive.
You can add the file extension, but it is not required. The only allowed extension is .tsp (see
Example 2).
Example 1
Saves the script referenced by the variable
test8
to nonvolatile memory.
Example 2
test8.save("/usb1/myScript.tsp")
Saves the script referenced by the variable
test8 to a file named myScript.tsp on your
USB flash drive.
Also see