7: TSP command reference Series 2600B System SourceMeter® Instrument
7-344 2600BS-901-01 Rev. C / August 2016
status.reset()
This function resets all bits in the status model.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
status.reset()
Details
This function clears all status data structure registers (enable, event, NTR, and PTR) to their default
values. For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status
register set contents (on page E-1) and Enable and transition registers (on page E-19).
Example
Resets the instrument status model.
Also see
Status model (on page 5-15, on page E-1)
status.standard.*
These attributes manage the standard event status register set of the status model.
Type TSP-Link accessible Affected by Where saved Default value
.event (R) Yes Status reset Not saved 0
Usage
standardRegister = status.standard.condition
standardRegister = status.standard.enable
standardRegister = status.standard.event
standardRegister = status.standard.ntr
standardRegister = status.standard.ptr
status.standard.enable = standardRegister
status.standard.ntr = standardRegister
status.standard.ptr = standardRegister
The status of the standard event status register; a zero (0) indicates no bits set (also
send 0 to clear all bits); other values indicate various bit settings