B: Calibration Series 2600B System SourceMeter® Instrument Referen
B-24 2600BS-901-01 Rev. C / August 2016
Calibration commands quick reference
The following table summarizes remote calibration commands. For a more complete description of
these commands, refer to the TSP command reference (on page 7-1
).
Calibration commands
Command** Description
smuX.cal.adjustdate = adjustDate
Set date when the adjustment was done.
Set calibration date (calDate of 0 indicates date not set).
Set date when calibration should be performed (calDue of 0
smuX.cal.password = "newPassword"
Change password to "newPassword".
smuX.cal.polarity = calPolarity
smuX.CAL_AUTO (automatic polarity).
smuX.CAL_NEGATIVE (negative polarity).
X
(positive polarity).
Load set of calibration constants:
smuX.CALSET_NOMINAL (nominal constants).
smuX.CALSET_FACTORY (factory constants).
smuX.CALSET_DEFAULT (normal constants).
X
(previous constants).
Store constants in nonvolatile memory as DEFAULT
calstate = smuX.cal.state
Request calibration state:
smuX.CALSTATE_CALIBRATING
smuX.CALSTATE_LOCKED
X
smuX.cal.unlock("password")
Unlock calibration (default password: KI0026XX)
smuX.measure.calibratei(range,
cp1Measured, cp1Reference,
cp2Measured, cp2Reference)
Adjust current measurement range calibration*:
±range (measurement range to adjust).
cp1Measured (Series 2600B measured value for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Measured (Series 2600B measured value for cal. point 2).
cp2Reference
(reference measurement for cal. point 2).
smuX.measure.calibratev(range,
cp1Measured, cp1Reference,
cp2Measured, cp2Reference)
Adjust voltage measurement range calibration*:
±range (measurement range to adjust).
cp1Measured (Series 2600B measured value for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Measured (Series 2600B measured value for cal. point 2).
cp2Reference (reference measurement for cal. point 2).
smuX.source.calibratei(range,
cp1Expected, cp1Reference,
cp2Expected, cp2Reference)
Adjust current source range calibration*:
±range (source range to adjust).
cp1Expected (source value programmed for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Expected (source value programmed for cal. point 2).
cp2Reference
(reference measurement for cal. point 2).
smuX.source.calibratev(range,
cp1Expected, cp1Reference,
cp2Expected, cp2Reference)
Adjust voltage source range calibration*:
±range (source range to adjust).
cp1Expected (source value programmed for cal. point 1).
cp1Reference (reference measurement for cal. point 1).
cp2Expected (source value programmed for cal. point 2).
cp2Reference (reference measurement for cal. point 2)