Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-318 2600BS-901-01 Rev. C / August 2016
Bit B15 B14 B13 B12 B11 B10 B9 B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal 32,768 16,384 8,192 4,096 2,048 1,024 512 256
Weights
(2
15
) (2
14
) (2
13
) (2
12
) (2
11
) (2
10
) (2
9
) (2
8
)
Example
status.operation.instrument.trigger_timer.trigger_overrun.enable =
status.operation.instrument.trigger_timer.trigger_overrun.TMR3
Uses a
constant to
set the timer
3 bit of the
operation
status trigger
timer overrun
enable
register.
Also see
Operation Status Registers (on page E-9)
status.operation.instrument.trigger_timer.* (on page 7-314)
status.operation.instrument.tsplink.*
This attribute contains the operation status TSP-Link summary register set. This attribute is not available on the
Models 2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
1024 (All bits set)
Usage
operationRegister = status.operation.instrument.tsplink.condition
operationRegister = status.operation.instrument.tsplink.enable
operationRegister = status.operation.instrument.tsplink.event
operationRegister = status.operation.instrument.tsplink.ntr
operationRegister = status.operation.instrument.tsplink.ptr
status.operation.instrument.tsplink.enable = operationRegister
status.operation.instrument.tsplink.ntr = operationRegister
status.operation.instrument.tsplink.ptr = operationRegister
operationRegister
The status of the operation status TSP-Link summary register; a zero (0) indicates
no bits set (also send 0 to clear all bits); the only valid value other than 0 is 1024
Details
These attributes are used to read or write to the operation status TSP-Link summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15.