System SourceMeter® Instrument Reference Manual Section 7: TSP command ref
2600BS-901-01 Rev. C / August 2016 7-219
smuX.measure.delayfactor
This attribute stores a multiplier to the delays that are used when smuX.measure.delay is set to
smuX.DELAY_AUTO.
Type TSP-Link accessible Affected by Where saved Default value
SMU reset
Recall setup
Usage
delayFactor = smuX.measure.delayfactor
smuX.measure.delayfactor = delayFactor
The delay factor multiplier
Source-measure unit (SMU) channel (for example, smua.measure.delayfactor
applies to SMU channel A)
Details
The delay factor is only applied when smuX.measure.delay = smuX.DELAY_AUTO.
This attribute can be set to a value less than 1 (for example, 0.5) to decrease the automatic delay.
This attribute can be set to a value greater than 1 (for example, 1.5 or 2.0) to increase the automatic
delay.
Setting this attribute to zero disables delays when smuX.measure.delay = smuX.DELAY_AUTO.
Example
smua.measure.delayfactor = 2.0
Doubles the measure delay for SMU channel A.
Also see
Measure auto delay (on page 2-82)
reset() (on page 7-173)
smuX.measure.delay (on page 7-218)
smuX.reset() (on page 7-232)
smuX.measure.filter.count
This command sets the number of measured readings that are required to yield one filtered measurement.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW) Yes Instrument reset
SMU reset
Saved setup 1
Usage
filterCount = smuX.measure.filter.count
smuX.measure.filter.count = filterCount
The number of readings required for each filtered measurement (1 to 100)
Source-measure unit (SMU) channel (for example,
smua.measure.filter.count applies to SMU channel A)