System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-295
Details
These attributes are used to read or write to the operation status calibration summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15.
For information about .condition, .enable, .event, .ntr, and .ptr registers, refer to
Status register set
contents (on page E-1) and Enable and transition registers (on page E-19). The individual bits of this
register are defined in the following table.
Bit Value Description
B1
status.operation.calibrating.SMUA
Set bit indicates that SMU A is unlocked for
calibration.
Bit B1 decimal value: 2
status.operation.calibrating.SMUB
This bit is only available on Models
2602B/2604B/2612B/2614B/2634B/2636B. Set bit
indicates that SMU B is unlocked for calibration.
Bit B2 decimal value: 4
Binary value: 0000 0100
Example
status.operation.calibrating.enable =
status.operation.calibrating.SMUA
Sets the SMUA bit of the operation
status calibration summary enable
register using a constant.
Also see
Operation Status Registers (on page E-9)
status.operation.* (on page 7-292)
status.operation.instrument.*
This attribute contains the operation status instrument summary register set.
Type TSP-Link accessible Affected by Where saved Default value
.ptr (RW) Yes Status reset Not saved Models 2601B/2611B/2635B:
31,746 (All bits set)
Models 2602B/2612B/2636B:
31,750 (All bits set)
Models 2604B/2614B/2634B: