7: TSP command reference Series 2600B System SourceMeter® Instrument
7-374 2600BS-901-01 Rev. C / August 2016
trigger.timer[N].clear()
This function clears the timer event detector and overrun indicator for the specified trigger timer number.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
trigger.timer[N].clear()
Trigger timer number (1 to 8)
Details
This command sets the timer event detector to the undetected state and resets the overrun indicator.
Example
Also see
trigger.timer[N].count (on page 7-374)
trigger.timer[N].count
This attribute sets the number of events to generate each time the timer generates a trigger event.
Type TSP-Link accessible Affected by Where saved Default value
Recall setup
Trigger timer N reset
Usage
count = trigger.timer[N].count
trigger.timer[N].count = count
Number of times to repeat the trigger (0 to 1,048,575)
Trigger timer number (1 to 8)
Details
If count is set to a number greater than 1, the timer automatically starts the next delay at the
expiration of the previous delay.
Set count to zero (0) to cause the timer to generate trigger events indefinitely.
Example
print(trigger.timer[1].count)
Read trigger count for timer number 1.
Also see
trigger.timer[N].clear() (on page 7-374)
trigger.timer[N].reset() (on page 7-378)