EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #361 background imageLoading...
Page #361 background image
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-22 2600BS-901-01 Rev. C / August 2016
Details
Assigning a value to this attribute enables or disables the storage of source values. Reading this
attribute returns the state of source value collection. This value can only be changed with an empty
buffer. Empty the buffer using the bufferVar.clear() function.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the
reading buffer is saved to nonvolatile memory.
When on, source values will be stored with readings in the buffer. This requires four extra bytes of
storage for each reading. Turning on additional collection items, such as source values (this attribute)
and timestamps, decreases the capacity of a dedicated reading buffer (for example,
smua.nvbuffer1), but does not change the capacity of a user-defined dynamically allocated buffer.
Example
smua.nvbuffer1.collectsourcevalues = 1
Include source values with readings for
dedicated reading buffer 1 (source-measure
unit (SMU) channel A).
Also see
bufferVar.clear() (on page 7-20)
Reading buffers (on page 3-6)
smuX.measure.overlappedY() (on page 7-225)
smuX.measure.Y() (on page 7-229)
smuX.nvbufferY (on page 7-231)
smuX.trigger.measure.Y() (on page 7-264)
bufferVar.collecttimestamps
This attribute sets whether or not timestamp values are stored with the readings in the buffer.
Type TSP-Link accessible Affected by Where saved Default value
Attribute (RW)
Yes
Not applicable
See Details
0 (disabled)
Usage
state = bufferVar.collecttimestamps
bufferVar.collecttimestamps = state
state
Timestamp value collection status; set to one of the following:
• 0: Timestamp value collection disabled (off)
•
1
: Timestamp value collection enabled (on)
bufferVar
The reading buffer. Can be a dynamically allocated user-defined buffer or a
dedicated reading buffer.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600B and is the answer not in the manual?

Keithley 2600B Specifications

General IconGeneral
BrandKeithley
Model2600B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals