EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #637 background imageLoading...
Page #637 background image
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-298 2600BS-901-01 Rev. C / August 2016
Bit B7 B6 B5 B4 B3 B2 B1 B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
) (2
6
) (2
5
) (2
4
) (2
3
) (2
2
) (2
1
) (2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
) (2
14
) (2
13
) (2
12
) (2
11
) (2
10
) (2
9
) (2
8
)
Example 1
operationRegister = status.operation.instrument.SMUA +
status.operation.instrument.TRGBLND
status.operation.instrument.enable = operationRegister
Sets bit B1 and bit B10 of the
operation status instrument
summary enable register using
constants.
Example 2
-- 1026 = binary 0000 0100 0000 0010
operationRegister = 1026
status.operation.instrument.enable = operationRegister
Sets bit B1 and bit B10 of the
operation status instrument
summary enable register using
a decimal value.
Also see
Operation Status Registers (on page E-9)
status.operation.* (on page 7-292)
Condition register sets of:
• status.operation.instrument.digio.* (on page 7-298)
• status.operation.instrument.lan.* (on page 7-302)
• status.operation.instrument.trigger_blender.* (on page 7-310)
• status.operation.instrument.trigger_timer.* (on page 7-314)
• status.operation.instrument.tsplink.* (on page 7-318)
status.operation.instrument.digio.*
This attribute contains the operation status digital I/O summary register set. This attribute is not available on the
Models 2604B/2614B/2634B.
Type TSP-Link accessible Affected by Where saved Default value
Attribute
- -
- -
- -
- -
.condition (R) Yes Not applicable Not saved Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
1024 (All bits set)