7: TSP command reference Series 2600B System SourceMeter® Instrument
7-328 2600BS-901-01 Rev. C / August 2016
Bit B7 B6 B5 B4 B3 B2 B1 B0
(2
) (2
) (2
) (2
) (2
) (2
) (2
) (2
)
Binary value 0/1 0/1 0/1 0/1 0/1 0/1 0/1 0/1
Weights
(2
) (2
) (2
) (2
) (2
) (2
) (2
) (2
)
Example
status.operation.trigger_overrun.SMUA +
status.operation.trigger_overrun.TRGBLND
status.operation.trigger_overrun.enable = operationRegister
Uses constants to set bit
B1 and bit B10 of the
operation status trigger
overrun summary enable
Also see
Operation Status Registers (on page E-9)
status.operation.* (on page 7-292)
status.operation.user.*
These attributes manage the operation status user register set of the status model.
Type TSP-Link accessible Affected by Where saved Default value
.enable (RW) Yes Status reset Not saved 0
Usage
operationRegister = status.operation.user.condition
operationRegister = status.operation.user.enable
operationRegister = status.operation.user.event
operationRegister = status.operation.user.ntr
operationRegister = status.operation.user.ptr
status.operation.user.condition = operationRegister
status.operation.user.enable = operationRegister
status.operation.user.ntr = operationRegister
status.operation.user.ptr = operationRegister
The status of the operation status user register; a zero (0) indicates no bits set
(also send 0 to clear all bits); other values indicate various bit settings