Series 2600B
System SourceMeter® Instrument Reference Manual Section 7: TSP command
reference
2600BS-901-01 Rev. C / August 2016 7-141
Also see
lan.lxidomain (on page 7-131)
lan.trigger[N].clear() (on page 7-141)
lan.trigger[N].mode (on page 7-145)
lan.trigger[N].overrun (on page 7-146)
lan.trigger[N].stimulus (on page 7-148)
lan.trigger[N].wait() (on page 7-150)
Understanding hardware value and pseudo line state (on page 3-51)
lan.trigger[N].clear()
This function clears the event detector for a trigger.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
lan.trigger[N].clear()
N
The LAN event number to clear (1 to 8)
Details
The trigger event detector enters the detected state when an event is detected. This function clears a
trigger event detector and discards the previous history of the trigger packet.
This function clears all overruns associated with this LAN trigger.
Example
lan.trigger[5].clear()
Clears the event detector with LAN packet 5.
Also see
lan.trigger[N].assert() (on page 7-140)
lan.trigger[N].overrun (on page 7-146)
lan.trigger[N].stimulus (on page 7-148)
lan.trigger[N].wait() (on page 7-150)