System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-55
Example
print(data)
Assume lines 2, 4, 6, and 8 are set high when
the I/O port is read.
Output:
1.70000e+02
Also see
digio.readbit() (on page 7-54)
digio.writebit() (on page 7-64)
digio.writeport() (on page 7-64)
Digital I/O port (on page 3-82)
digio.trigger[N].assert()
This function asserts a trigger on one of the digital I/O lines. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
digio.trigger[N].assert()
Digital I/O trigger line (1 to 14)
Details
The set pulsewidth determines how long the trigger is asserted.
Example
digio.trigger[2].assert()
Asserts a trigger on digital I/O line 2.
Also see
digio.trigger[N].pulsewidth (on page 7-58)
digio.trigger[N].clear()
This function clears the trigger event on a digital I/O line. This function is not available on the Models
2604B/2614B/2634B.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
digio.trigger[N].clear()
Digital I/O trigger line (1 to 14)