7: TSP command reference Series 2600B System SourceMeter® Instrument
7-274 2600BS-901-01 Rev. C / August 2016
smuX.trigger.SOURCE_COMPLETE_EVENT_ID
This constant contains the source complete event number.
Type TSP-Link accessible Affected by Where saved Default value
Usage
eventID = smuX.trigger.SOURCE_COMPLETE_EVENT_ID
The source action complete event number
Source-measure unit (SMU) channel (for example,
smua.trigger.SOURCE_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
source complete events from this source-measure unit (SMU).
Also see
Triggering (on page 3-32)
smuX.trigger.SWEEP_COMPLETE_EVENT_ID
This constant contains the sweep complete event number.
Type TSP-Link accessible Affected by Where saved Default value
Constant Yes
Usage
eventID = smuX.trigger.SWEEP_COMPLETE_EVENT_ID
The sweep complete event number
Source-measure unit (SMU) channel (for example,
smua.trigger.SWEEP_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
sweep complete events from this SMU.
Also see
Triggering (on page 3-32)
smuX.trigger.SWEEPING_EVENT_ID
This constant contains the sweeping event number.
Type TSP-Link accessible Affected by Where saved Default value