EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #495 background imageLoading...
Page #495 background image
Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-156 2600BS-901-01 Rev. C / August 2016
Usage
prompting = localnode.prompts4882
localnode.prompts4882 = prompting
prompting
IEEE Std 488.2 prompting mode:
• Disable prompting: 0
•
Enable prompting: 1
Details
When this attribute is enabled, the IEEE Std 488.2 common commands generate prompts if
prompting is enabled with the localnode.prompts attribute. If localnode.prompts4882 is
enabled, limit the number of *trg commands sent to a running script to 50 regardless of the setting
of the localnode.prompts attribute.
When this attribute is disabled, IEEE Std 488.2 common commands will not generate prompts. When
using the *trg command with a script that executes trigger.wait() repeatedly, disable
prompting to avoid problems associated with the command interface input queue filling.
When using this command from a remote node, localnode should be replaced with the node
reference, for example:
node[5].prompts4882
Example
localnode.prompts4882 = 0
Disables IEEE Std 488.2 common command prompting.
Also see
localnode.prompts (on page 7-154)
localnode.reset()
This function resets the local node instrument.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
localnode.reset()
Details
If you want to reset a specific instrument or a subordinate node, use the node[X].reset()
command.
A local node reset includes:
• Source-measure unit (SMU) attributes affected by a SMU reset are reset
• Other settings are restored back to factory default settings
A localnode.reset() is different than a reset() because reset() resets the entire system.
When using this command from a remote node, localnode should be replaced with the node
reference, for example node[5].reset().