7: TSP command reference Series 2600B System SourceMeter® Instrument
7-156 2600BS-901-01 Rev. C / August 2016
Usage
prompting = localnode.prompts4882
localnode.prompts4882 = prompting
IEEE Std 488.2 prompting mode:
• Disable prompting: 0
•
Details
When this attribute is enabled, the IEEE Std 488.2 common commands generate prompts if
prompting is enabled with the localnode.prompts attribute. If localnode.prompts4882 is
enabled, limit the number of *trg commands sent to a running script to 50 regardless of the setting
of the localnode.prompts attribute.
When this attribute is disabled, IEEE Std 488.2 common commands will not generate prompts. When
using the *trg command with a script that executes trigger.wait() repeatedly, disable
prompting to avoid problems associated with the command interface input queue filling.
When using this command from a remote node, localnode should be replaced with the node
reference, for example:
node[5].prompts4882
Example
localnode.prompts4882 = 0
Disables IEEE Std 488.2 common command prompting.
Also see
localnode.prompts (on page 7-154)
localnode.reset()
This function resets the local node instrument.
Type TSP-Link accessible
Affected by Where saved Default value
Function Yes
Usage
localnode.reset()
Details
If you want to reset a specific instrument or a subordinate node, use the node[X].reset()
command.
A local node reset includes:
• Source-measure unit (SMU) attributes affected by a SMU reset are reset
• Other settings are restored back to factory default settings
A localnode.reset() is different than a reset() because reset() resets the entire system.
When using this command from a remote node, localnode should be replaced with the node
reference, for example node[5].reset().