EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #856 background imageLoading...
Page #856 background image
Series 2600B
System SourceMeter® Instrument Reference Manual Appendix E: Status m
odel
2600BS-901-01 Rev. C / August 2016 E-15
When using a numeric parameter, registers are programmed by including the appropriate mask
value. For example:
*ese 1169
status.standard.enable = 1169
To convert from decimal to binary, use the information shown in the above figure. For example, to set
bits B0, B4, B7, and B10, a decimal value of 1169 would be used for the mask parameter (1169 = 1 +
16 + 128 + 1024).
Reading registers
Any register in the status structure can be read either by sending the common command query
(where applicable), or by including the script command for that register in either the print() or
print(tostring()) command. The print() command outputs a numeric value; the
print(tostring()) command outputs the string equivalent. For example, any of the following
commands requests the Service Request Enable Register value:
*SRE?
print(tostring(status.request_enable))
print(status.request_enable)
The response message will be a decimal value that indicates which bits in the register are set. That
value can be converted to its binary equivalent using the information in
Programming enable and
transition registers (on page E-14). For example, for a decimal value of 37 (binary value of 100101),
bits B5, B2, and B0 are set.
Status byte and service request (SRQ)
Service requests (SRQs) allow an instrument to indicate that it needs attention or that some event
has occurred. When the controller receives an SRQ, it allows the controller to interrupt tasks to
perform other tasks in order to address the request for service.
For example, you might program your instrument to send an SRQ when:
• All instrument operations are complete
• An instrument error occurs
• A specific operation has occurred
Two 8-bit registers control service requests, the Status Byte Register and the Service Request
Enable Register. The Status Byte Register (on page E-16
) topic describes the structure of these
registers.
Service requests affect GPIB, USB, and VXI-11 connections. On a GPIB connection, the SRQ line is
asserted. On a VXI-11 or USB connection, an SRQ event is generated.