System SourceMeter® Instrument Reference Manual Section 7:
2600BS-901-01 Rev. C / August 2016 7-45
ConfigPulseVMeasureISweepLin()
This KIPulse factory script function configures a voltage pulse train with a current measurement at each point.
Type TSP-Link accessible
Affected by Where saved Default value
Usage
f, msg = ConfigPulseVMeasureISweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout, sync_in_abort)
f, msg = ConfigPulseVMeasureISweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out, sync_in_timeout)
f, msg = ConfigPulseVMeasureISweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in, sync_out)
f, msg = ConfigPulseVMeasureISweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag, sync_in)
f, msg = ConfigPulseVMeasureISweepLin(smu, bias, start, stop, limit, ton, toff,
points, buffer, tag)
A Boolean flag; this flag is true when the pulse was successfully configured, false when
A string message; if the f flag is false, msg contains an error message; if it is true, msg
contains a string indicating successful configuration
System SourceMeter
instrument channel (for example,
refers to SMU channel A)
Bias level in volts
start
Pulse sweep start level in volts
Pulse sweep stop level in volts
Current limit (for example, compliance) in amperes
Pulse on time in seconds
Pulse off time in seconds
Number of pulse-measure cycles
Reading buffer where pulsed measurements will be stored; if this is nil when the function is
called, no measurements will be made when the pulse train is initiated
Numeric identifier to be assigned to the defined pulse train
Defines a digital I/O trigger input line; if programmed, the pulse train will wait for a trigger input
before executing each pulse
Defines a digital I/O trigger output line; if programmed, the pulse train will generate a trigger
output immediately before the start of
Specifies the length of time (in seconds) to wait for input trigger; default value is 10 s
Specifies whether or not to abort pulse if input trigger is not received; if pulse aborts because
of a missed trigger, a timer timeout message is returned; true or false
Details
Data for pulsed current measurements are stored in the reading buffer specified by the buffer input
parameter.
This function configures a linear pulsed voltage sweep with a current measurement at each point.
Measurements are made at the end of the ton time.