7: TSP command reference Series 2600B System SourceMeter® Instrument
7-46 2600BS-901-01 Rev. C / August 2016
The magnitude of the first pulse will be start volts; the magnitude of the last pulse will be stop
volts. The magnitude of each pulse in between will be step volts larger than the previous pulse,
where:
step = (stop - start) / (points - 1)
This function does not cause the specified smu to output a pulse train. It does check to see if all the
pulse dimensions can be achieved, and if they can, assigns the indicated tag or index to the pulse
train.
The InitiatePulseTest(tag) and InitiatePulseTestDual(tag1, tag2) functions are
used to initiate a pulse train assigned to a valid tag.
Figure 129: ConfigPulseVMeasureISweepLin()
Example
ConfigPulseVMeasureISweepLin(smua, 0, 1,
10, 1, 10e-3, 20e-3, 16,
smua.nvbuffer1, 4)
Set up a pulsed sweep that uses System
SourceMeter
®
instrument channel A. The pulsed
sweep starts at 1 V, ends at 10 V, and returns to a
0 V bias level between pulses. Each pulsed step is on
for 10 ms, and then at the bias level for 20 ms.
The current limit is 1 A during the entire pulsed
sweep. The pulse train is comprised of 16 pulsed
steps, and the pulse train is assigned a tag index of
Also see
InitiatePulseTest() (on page 7-112)
InitiatePulseTestDual() (on page 7-114)
KIPulse factory script (on page 5-22)