Section
3: Functions and features Series 2600B System SourceMeter® Instrument
Reference Manual
3-64 2600BS-901-01 Rev. C / August 2016
Figure 81: Synchronous input trigger
Output characteristics:
• In addition to trigger events from other trigger objects, the digio.trigger[N].assert() and
tsplink.trigger[N].assert() functions generate a low pulse for the programmed pulse
duration if the line is latched low, a falling edge does not occur.
• A normal falling edge pulse generates when the internal drive is not latched low and the
digio.trigger[N].assert() and tsplink.trigger[N].assert() functions are issued.
• To mirror the SynchronousA trigger mode, set the pulse width to 1 µs or any small nonzero value.
• Action overruns are disabled.
Figure 82: Synchronous output trigger