EasyManuals Logo
Home>Keithley>Measuring Instruments>2600B

Keithley 2600B User Manual

Keithley 2600B
897 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #230 background imageLoading...
Page #230 background image
Section
4: Theory of operation Series 2600B System SourceMeter® Instrument
Reference Manual
4-6 2600BS-901-01 Rev. C / August 2016
Continuous power operating boundaries
The general operating boundaries for Model 2601B/2602B/2604B continuous power output are
shown in the following figure (for derating factors, see the General power equation (on page 4-4
)
described earlier in this section). In this drawing, the current (600 mA, 1 A, 2.2 A, and 3 A) and the
voltage (6 V and 40 V) magnitudes are nominal values. Also note that the boundaries are not drawn
to scale.
Figure 90: Model 2601B/2602B/2604B continuous power operating boundaries
The general operating boundaries for Model 2611B/2612B/2614B/2634B/2635B/2636B continuous
power output are shown in the following figure (for derating factors, see the General power equation
(on page 4-4) described earlier in this section). In this drawing, the current (100 mA and 1.5 A) and
voltage (20 V and 200 V) magnitudes are nominal values. Also note that the boundaries are not
drawn to scale.
Figure 91: Model 2611B/2612B/2614B/2634B/2635B/2636B continuous power operating
boundaries

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600B and is the answer not in the manual?

Keithley 2600B Specifications

General IconGeneral
BrandKeithley
Model2600B
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals