Section
7: TSP command reference Series 2600B System SourceMeter® Instrument
Reference Manual
7-48 2600BS-901-01 Rev. C / August 2016
Details
Data for pulsed current measurements are stored in the reading buffer specified by the buffer input
parameter.
This function configures a logarithmic pulsed voltage sweep with a current measurement at each
point. Measurements are made at the end of the ton time.
The magnitude of the first pulse will be start volts; the magnitude of the last pulse will be stop
volts. The magnitude of each pulse in between will be LogStep
n
volts larger than the previous pulse,
where:
LogStepSize = (log10(stop) - log10(start)) / (points -1)
LogStep
n
= (n - 1) * (LogStepSize), where n = [2, points]
SourceStepLevel
n
= antilog(LogStep
n
) * start
This function does not cause the specified smu to output a pulse train. It does check to see if all the
pulse dimensions can be achieved, and if they can, assigns the indicated tag or index to the pulse
train. The InitiatePulseTest(tag) and InitiatePulseTestDual(tag1, tag2) functions
are used to initiate a pulse train assigned to a valid tag.
Figure 130: ConfigPulseVMeasureISweepLog()

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Keithley 2600B and is the answer not in the manual?

Keithley 2600B Specifications

General IconGeneral
TypeSource Measure Unit (SMU)
Operating Temperature0°C to 50°C
Storage Temperature-40°C to 70°C
Model2600B
Channels1 or 2
Resolution (Voltage)100nV
Resolution (Current)10fA
Output Power20W per channel
Maximum Power20W per channel
Resolution6.5 digits
ConnectivityGPIB, USB, Ethernet

Related product manuals