7: TSP command reference Series 2600B System SourceMeter® Instrument
7-284 2600BS-901-01 Rev. C / August 2016
Usage
measurementRegister = status.measurement.instrument.smuX.condition
measurementRegister = status.measurement.instrument.smuX.enable
measurementRegister = status.measurement.instrument.smuX.event
measurementRegister = status.measurement.instrument.smuX.ntr
measurementRegister = status.measurement.instrument.smuX.ptr
status.measurement.instrument.smuX.enable = measurementRegister
status.measurement.instrument.smuX.ntr = measurementRegister
status.measurement.instrument.smuX.ptr = measurementRegister
The status of the instrument measurement status SMU X summary register; a
zero (0) indicates no bits set (also send 0 to clear all bits); other values indicate
Source-measure unit (SMU) channel (for example
status.measurement.instrument.smua.enable applies to SMU
Details
These attributes are used to read or write to the measurement event SMU X summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15. For example, assume the value 257 is returned for the enable register. The binary
equivalent is 0000 0001 0000 0001. This value indicates that bit B0 (VLMT) and bit B8 (BAV) are set.