EasyManuals Logo
Home>R&S>Control Unit>OSP

R&S OSP User Manual

R&S OSP
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
R&S OSP Remote Command Reference
Operating Manual 1505.3896.12 - 14 165
'<file_name>'
String parameter to specify the name of the file to be removed.
Example
MMEM:DEL "my.pathconfigs"
Removes file my.pathconfigs from the flash memory.
Characteristics
Firmware version V1.0
SCPI-confirmed. No query.
7.4.5 DIAGnostic Commands
The purpose of the DIAGNOSTIC subsystem is to provide instrument service and
diagnostic commands. All commands in this subsystem are device-specific.
DIAGnostic:SERVice:STESt:RESult?
Starts an instrument internal self test routine and returns information about any error
that is detected.
Example - Query
DIAG:SERV:STES:RES?
Response: “Passed”
or, for example if an error is detected:
“Failed, OSP-B101-F01A11, Error XY occurred.”
Characteristics
Firmware version V1.0
Device-specific. Query only.
DIAGnostic:SERVice:HWINfo?
Returns information about all components (mainboards and modules) that are part of
the complete system of one or several R&S OSP instruments.
The returned component information comprises:
location ID
name
serial number
part number
hardware code
product index.
The response is a string in following format:
”<location>|<name>|<sn_nbr>|<part_nbr>|<hardware_code>|<product_index>“
Example Query
DIAG:SERV:HWIN?
Response:
“F01|OSPMAINBOARD|100123|1505.3050.04|0|03.00”,
“F01A11|OSP-B101|100001|1505.3250.02|0|01.00”,
“F01A12|OSP-B102|100002|1505.3260.02|0|01.00”,

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S OSP and is the answer not in the manual?

R&S OSP Specifications

General IconGeneral
BrandR&S
ModelOSP
CategoryControl Unit
LanguageEnglish