Section 8.0
MAN 06–0017H
ZIEHM QUANTUM User’s Manual Page 114 of 220
double tap the image on the DeskView to have it become the active image.
3. Press the W/L button. The system will open the adjustment buttons. Use the function
buttons to adjust the image contrast and brightness level to a relative image level. The
image can then be stored to the hard drive, transferred to the right monitor, USB device or
DICOM server.
4. Press the Reset W/L button to return the image to its original contrast and brightness
levels.
5. To save the new window and level values as the new defaults for image display press the
Save W/L button.
6. To exit the Window Level function, press the W/L button again or one of the other post
processing buttons to close W/L and open the new function.
8.20.2. INVERT
The INVERT Video function can be used with both stored images and real-time flu-
oroscopy.
Press the INVERT Video
utton to reverse the light and dark areas of the image and
create an image that resembles a film negative.
The effect can be changed latter by recalling the image and pressing the INVERT
button again.
When you press store the image will be stored with the INVERT image attribute so the
image will be recalled with the Black and White levels inverted.
Fig 8.2: INVERT Touch Screen Display