19-4 Passport 12/Passport 8 Operator’s Manual
19.4.2 Reviewing Events
The monitor saves the events in real time. You can review these events.
In the [Review] menu, select [Events] to access the following window.
You can review the parameter alarm events, arrhythmia alarm events and manual events. When an event occurs, the
system stores all the measurement numerics at the event triggering time and related waveforms 4 seconds, 8 seconds,
or 16 seconds before and after the event triggering time, as per the setting of recording length.
NOTE
Pausing or switching off alarms will not be recorded as events. The time of these operations will not be
recorded in the system log.
Earlier-recorded events might be overwritten by later ones if the events log reaches capacity.
In this window:
Select [Event] to view the desired events.
Selecting [Level] to view the desired events according to the alarm priority .
After selecting the desired event, you can select [Details] to access the following window. In this window, the waveform
area displays the waveforms related to the event, and the parameter area displays the parameter values happened at the
event trigger time.
NOTE
A total loss of power has no impact on the saved events.