C-16 Passport 12/Passport 8 Operator’s Manual
C.3 User Maintenance Items
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Changing Bed No. No Yes Protected
Atmospheric Pressure No Yes 760 mmHg
Height Unit No Yes cm
Weight Unit No Yes kg
ST Unit No Yes mV
Press. Unit No Yes mmHg
CVP Unit No Yes cmH
2
O
CO
2
Unit No Yes mmHg
O
2
Unit No Yes %
Temp Unit No Yes °C
Network Type No Yes LAN
Address Type No Yes Manual
Select CMS No Yes On
ADT Query No Yes On
Latching Alarms No Yes Off
Alarm Pause Time No Yes 2 min
Minimum Alarm Volume No Yes
General, ICU, NICU, CCU:
OR:
2
1
Alarm Sound No Yes ISO
High Alarm Interval (ISO mode) No Yes 10
Medium (ISO mode) No Yes 20
Low Alarm Interval (ISO mode) No Yes 20
Reset Other Bed’s Alarms No Yes Off
Alarm Reset By Other Bed No Yes On
Other Bed Disconnection Alm No Yes On
Reminder Interval No Yes 1 min
Reminder Tone No Yes On
ECGLeadOff Lev. No Yes Low
SpO
2
SensorOff Lev. No Yes Low
IBPSensorOff Lev. No Yes Med
Lethal Arrh. OFF No Yes Disable
Extended Arrh. No Yes Enable
Alarm Light on Alarm Reset No Yes On
Alarm Delay No Yes 6 s
ST Alarm Delay No Yes 30 s
Wave Line No Yes Med
DIAP Baud Rate No Yes 9600
ECG Standard No Yes AHA
Notch Freq. No Yes 50 Hz
Data Transfer Method No Yes Module