Passport 12/Passport 8 Operator’s Manual 4-3
4.4 Associating Patient Information
After associating patient information with the HIS, the monitor automatically updates the patient information if the
corresponding HIS information changes. The monitor can associate the patient’s MRN (Medical Record Number), first
name, last name, date of birth, and gender with the HIS.
NOTE
A keyword takes effect only when being defined in the eGateway. Refer to eGateway Integration Manager
Installation Guide (P/N: 046-002447-00) for details.
The monitor displays corresponding patient information only when all the keywords have been entered.
4.5 Editing Patient Information
To edit the patient information after a patient has been admitted, when the patient information is incomplete, or when
you want to change the patient information:
1. Select the [Patient Setup] QuickKey, or [Main Menu]→[Patient Setup >>].
2. Select [Patient Demographics] and then make the required changes.
3. Select [Ok].
4.6 Discharging a Patient
To discharge a patient:
1. Select the [Patient Setup] QuickKey, or [Main Menu]→[Patient Setup >>].
2. Select [Discharge Patient]. In the pop-up menu, you can either:
Select [Ok] to discharge the current patient, or
Select [Standby] then [Ok]. The monitor enters the Standby mode after discharging the current patient, or
Select [Cancel] to exit without discharging the patient.
NOTE
After discharging an admitted patient, you can review the history data of this patient by selecting [Main
Menu]→[Patient Data >>]→[History Data >>]. When the SD card is full, the oldest patient data will be
deleted automatically. Therefore, you cannot always view all the history data.
4.7 Transferring Patient Data
You can transfer patient data between monitors with a USB drive without re-entering the patient demographic
information. Transferring of patient data enables you to review the patient’s condition history. The patient data that can
be transferred includes: patient demographics, trend data, alarm events and parameter alarm limits.
From the [User Maintenance] menu, select [Others >>]. In the pop-up menu, you can set [Transferred Data Length].
The default is [4 h]. You can also set [Data Transfer Method]. The default is [Off].