7-6 Passport 12/Passport 8 Operators Manual
7.5 Understanding the Alarm Setup Menu
Select the [Alarm Setup] QuickKey or [Main Menu][Alarm Setup >>] to enter the [Alarm Setup], where you can:
Set alarm properties for all parameters.
Change ST alarm settings.
Change arrhythmia alarm settings.
Set the threshold for some arrhythmia alarms.
Change other settings.
NOTE
The On/Off column is only visible when used with a CMS. This column is not visible with Panorama CMS.
Refer to the chapter 8 Monitoring ECG for how to change ST alarm settings, how to change arrhythmia alarm settings
and how to set the threshold for some arrhythmia alarms.
7.5.1 Setting Alarm Properties for All Parameters
In the [Main Menu], select [Alarm Setup >>][Parameters]. You can review and set alarm limits, alarm switches, alarm
level and alarm recordings for all parameters.
When a measurement alarm occurs, automatic recording of all the measurement numerics and related waveforms is
possible when the measurements [On/Off] and [Record] are set on.

Table of Contents

Other manuals for Mindray Passport 8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 8 and is the answer not in the manual?

Mindray Passport 8 Specifications

General IconGeneral
BrandMindray
ModelPassport 8
CategoryMedical Equipment
LanguageEnglish

Related product manuals