EasyManuals Logo
Home>Mindray>Medical Equipment>Passport 8

Mindray Passport 8 Operator's Manual

Mindray Passport 8
284 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
Passport 12/Passport 8 Operators Manual 26-3
Check/Maintenance Item Recommended Frequency
Performance test Once a year or if the battery run time reduced significantly.
26.3 Checking Monitor and Module Information
To view the information about system start time, selftest, etc., select [Main Menu][Maintenance >>][Monitor
Information >>]. You can print out the information for the convenience of troubleshooting. The information will not be
saved during shut down.
You can also view the information about the monitor configuration and system software version by selecting [Main
Menu][Maintenance >>][Software Version >>].
26.4 ECG Verification
The ECG signal may be inaccurate due to hardware or software problems. As a result, the ECG wave amplitude becomes
greater or smaller. In that case, you need to verify the ECG module.
1. Select the ECG parameter window or waveform area[Filter][Diagnostic].
2. Select [Main Menu][Maintenance >>][User Maintenance >>]enter the required password[Module
Maintenance >>][Calibrate ECG]. A square wave appears on the screen and the message [ECG Calibrating] is
displayed.
3. Compare the amplitude of the square wave with the wave scale. The difference should be within 5%.
4. After the calibration is completed, select [Stop Calibrating ECG]
You can print the square wave and wave scale and then measure the difference between them if necessary. If the
difference exceeds 5%, contact your service personnel.
26.5 NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. It is required at least once a year or when you
doubt the measured NIBP. If the test failed, corresponding prompt messages will be given. If no message is displayed, it
means no leakage is detected.
Tools required:
An adult cuff
An air tubing
A correct sized cylinder
Follow this procedure to perform the leakage test:
1. Set the patient category to [Adu].
2. Connect the cuff to the NIBP connector on the monitor.
3. Wrap the cuff around the cylinder as shown below.

Table of Contents

Other manuals for Mindray Passport 8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 8 and is the answer not in the manual?

Mindray Passport 8 Specifications

General IconGeneral
BrandMindray
ModelPassport 8
CategoryMedical Equipment
LanguageEnglish

Related product manuals