EasyManuals Logo
Home>Mindray>Medical Equipment>Passport 8

Mindray Passport 8 Operator's Manual

Mindray Passport 8
284 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #262 background imageLoading...
Page #262 background image
C-12 Passport 12/Passport 8 Operator’s Manual
Item Name
Configurable
Default
In Config Mode In Monitor Mode
FiN
2
O Low Yes Yes 0
EtHal/Enf/Iso High Yes Yes 3.0
EtHal/Enf/Iso Low Yes Yes 0.0
FiHal/Enf/Iso High Yes Yes 2.0
FiHal/Enf/Iso Low Yes Yes 0.0
EtSev High Yes Yes 6.0
EtSev Low Yes Yes 0.0
FiSev High Yes Yes 5.0
FiSev Low Yes Yes 0.0
EtDes High Yes Yes 8.0
EtDes Low Yes Yes 0.0
FiDes High Yes Yes 6.0
FiDes Low Yes Yes 0.0
C.2 Routine Configuration
C.2.1 Alarm
Item Name
Configurable
Default
In Config Mode In Monitor Mode
Alm Volume Yes Yes
General: 2
OR:
ICU, NICU, CCU:
2
1
2
Reminder Vol Yes Yes Low
Recording Length Yes Yes 16 s
Apnea Delay Yes Yes
Adu, Ped:
Neo:
20s
15 s
Alarm Delay Yes Yes 6 s
ST Alarm Delay Yes Yes 30 s