Passport 12/Passport 8 Operator’s Manual 3-11
3.12.2 Accessing the Parameters Menu
Select [Parameters >>] from the [Main Menu] or select corresponding parameter area or waveform area to access a
parameter setup menu.
3.13 Operating Modes
Your monitor has different operating modes. Some are password protected. This section lists the major operating
modes.
3.13.1 Monitoring Mode
This is the normal, everyday working mode that you use for monitoring patients. Your monitor automatically enters
the Monitoring mode after being turned on.
3.13.2 Night Mode
To avoid disturbing the patient, Night mode may be used.
To activate the Night mode:
1. Select the [Night Mode] QuickKey or [Main Menu]→[Screen Setup >>]→[Night Mode >>].
2. In the pop-up menu, set the desired brightness, alarm volume, QRS volume, key volume, NIBP end tone, or
whether to stop NIBP measurements or not. When [Stop NIBP] is selected, all the NIBP measurements terminate
after entering the Night mode.
3. Select the [Enter Night Mode] button.
To cancel the Night mode:
1. Select the [Night Mode] QuickKey or [Main Menu]→[Screen Setup >>]→[Night Mode >>].
2. Select [Ok] in the pop-up.
CAUTION
Before entering Night mode, confirm the brightness settings, alarm volume, QRS volume, and key volume.
Consider the potential risk when the setting value is low.
3.13.3 Privacy Mode
Privacy mode is only available when an admitted patient at a monitor is also monitored by the CMS.
To activate the Privacy mode, select [Main Menu]→[Screen Setup >>]→[Privacy Mode].