Passport 12/Passport 8 Operators Manual D-1
D Alarm Messages
This chapter lists only the most important physiological and technical alarm messages. Some messages appearing on
your monitor may not be included.
In this chapter:
The “I” column indicates how indications of technical alarms perform after the alarm system is reset: “A” means that
some technical alarms are cleared; “B” indicates that some technical alarms are changed to the prompt messages;
and “C” indicates that a “” appears before the alarm message, appears in the alarm symbol area, and the
indication of the alarm lamp depends on the alarm light setting. Refer to section 7.8 Resetting Alarms for details.
The “Level” field indicates the alarm level: H means high, M means medium and L means low. “*” means the alarm
level is user-adjustable.
XX represents a measurement or parameter label, such as ECG, NIBP, HR, ST-I, PVCs, RR, SpO
2
, PR, etc.
In the “Cause and Solution column, corresponding solutions are given instructing you to troubleshoot problems. If the
problem persists, contact your service personnel.
D.1 Physiological Alarm Messages
Measurement Alarm messages L Cause and solution
XX
XX Too High M* XX value has risen above the high alarm limit or fallen below the low
alarm limit. Check the patient’s condition and check if the patient
category and alarm limit settings are correct.
XX Too Low M*
ECG
ECG Weak Signal H
The ECG signal is too weak for the monitor to perform ECG analysis.
Check the patient’s condition and the ECG connections.
Asystole H
Arrhythmia has occurred to the patient. Check the patients
condition and the ECG connections.
VFib/VTac H
Vtac H
Vent. Brady H
Extreme Tachy H
Extreme Brady H
R on T M*
Run PVCs M*
Couplet M*
PVCs/min M*
Bigeminy M*
Trigeminy M*

Table of Contents

Other manuals for Mindray Passport 8

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Passport 8 and is the answer not in the manual?

Mindray Passport 8 Specifications

General IconGeneral
BrandMindray
ModelPassport 8
CategoryMedical Equipment
LanguageEnglish

Related product manuals