25-2 Passport 12/Passport 8 Operator’s Manual
25.2 Cleaning
Your equipment should be cleaned on a regular basis. Before cleaning the equipment, consult your hospital’s
regulations for cleaning the equipment.
The following table lists approved cleaning agents:
Product Name Product Type Ingredients
Sodium hypochlorite bleach Liquid Sodium hypochlorite 10%
Hydrogen peroxide Liquid Hydrogen peroxide 3%
Isopropanol Liquid Isopropanol 70%
Rely+On
TM
Virkon
Ⓡ
Powder Used as 1% solution*
Biocidal active:
Pentapotassium bis
(peroxymonosulphate) bis (sulphate)
(500g/kg)
Contains dipotassium peroxodisulphate.
1-Propanol Liquid 1-Propanol 50%
Perform
Ⓡ
Powder Used as 1% solution*
100g contains 45g Pentapotassium bis
(peroxymonosulphate) bis (sulphate).
5 % to 15% anionic
surfactants;<5% non-ionic
surfactants; <5% soap and
phosphonates, perfumes
Note: The “*” symbol indicates a cleaning solution should be prepared according to the concentration directed by the
accompanying instructions for use.
To clean your equipment, follow these rules:
1. Shut down the monitor and disconnect it from AC power.
2. Clean the display screen using a soft, clean cloth dampened with a cleaning agent.
3. Clean the exterior surface of the equipment using a soft cloth dampened with a cleaning agent.
4. Wipe off all the cleaning solution with a dry cloth after cleaning if necessary.
5. Dry your equipment in a ventilated, cool place.