Passport 12/Passport 8 Operator’s Manual 3-1
3 Basic Operations
3.1 Installation
WARNING
Connect only approved devices to this equipment. Devices connected to the equipment must meet the
requirements of the applicable IEC standards (e.g. IEC 60950 safety standards for information technology
equipment and IEC 60601-1 safety standards for medical electrical equipment). The system configuration
must meet the requirements of the IEC 60601-1 medical electrical systems standard. Any personnel who
connect devices to the equipment’s signal input/output port is responsible for providing evidence that the
safety certification of the devices has been performed in accordance to the IEC 60601-1. If you have any
questions, please contact Mindray.
If it is not evident from the equipment specifications whether a particular combination with other devices is
hazardous, for example, due to summation of leakage currents, please consult the manufacturers or else an
expert in the field, to ensure the necessary safety of patients and all devices concerned will not be impaired
by the proposed combination.
CAUTION
The equipment should be installed by authorized Mindray personnel.
NOTE
The software equipment copyright is solely owned by Mindray. No organization or individual shall resort to
modifying, copying, or exchanging it or to any other infringement on it in any form or by any means without
due permission.
Many settings in the patient monitor are password protected. It is recommended to change the default
passwords and keep the passwords safe. Passwords should only be changed by authorized personnel.
Contact your department manager or biomedical engineering department for the passwords used at your
facility.