1-225
IM 04L51B01-01EN
Configuring the GX/GP and Viewing the Settings
1
Note
If the advanced security function (/AS option) is in use, passwords are not loaded except for
monitor users. All administrator and user passwords are set to their defaults.
See “Reusing Setting Parameters” section 1.3.3, “User Levels,” in the Advanced Security
Function (/AS) User’s Manual (IM 04L51B01-05EN).
Control settings
Set this to On to load control settings.
Setup items that are loaded
• Control settings
• Program pattern settings
• Control event action
IP address
Set this to On to load only the IP address, DHCP, and DNS settings.
Other
Set this to On to load settings other than security, IP address, and control settings.
Communication(Serversettings)
Set this to On when you want to load the server list settings, server encryption settings, and
channel conversion setting of DARWIN compatible communication.
Calibration correction settings
Set this to On to load calibration correction settings and calibration reminder settings (/AH).
Instruments tags, etc.
Set this to On to load instrument tag settings.
Procedure
Loading All Stings
1
Tap Media kind, set the medium.
2
Tap Select file, set the file.
3
Tap All settings, set to On.
4
Tap Execute.
The setting parameters that you specied are loaded.
5
Tap Exit.
The GX/GP returns to the original screen.
Operation complete
Loading Specific Items
1
Tap Media kind, set the medium.
2
Tap Select file, set the file.
3
Tap All settings, set to Off.
4
Set the items that you want to load to On.
5
Tap Execute.
The setting parameters that you specied are loaded.
6
Tap Exit.
The GX/GP returns to the original screen.
Operation complete
1.25 Loading Settings