1-224
IM 04L51B01-01EN
1.25 Loading Settings
Load and delete setting parameters, report templates, trusted certificates, custom displays (/
CG option), program patterns (/PG option) (release number 4 and later), and scale images.
1.25.1 Loading Setting Parameters
Path
GX/GP: MENU key > Browse tab > Save load > Menu Load settings > Setting
parameters
Web application: Data tab > Save/Load > Setting parameters
Description
Setup Item Selectable Range or Options Default Value
Media kind SD, USB
1
Select file — —
1 Default value of Media kind
Media status Default Value
Non Media is not recognized
SD memory card SD
SD memory card, USB flash memory (/UH
option)
SD
USB meory (/UH option) USB
Media kind
Set the type of media.
Select file
Set the setting parameter file to load.
Setup Item Selectable Range or Options Default Value
All settings Off, On On
All settings
Set this to On to load all settings. Set this to Off to load security, control settings, IP address,
and other settings separately.
Setup Item Selectable Range or Options Default Value
Security
1
Off, On On
Control settings
3
Off, On On
IP address
1
Off, On On
Other
1
Off, On On
Communication(Server settings)
2
Off, On On
Calibration correction settings
2
Off, On On
Instruments tags, etc.
2
Off, On On
1 Appears when All settings is set to Off.
2 Appears when Other is set to On. Release number 3 and later.
3 Appears when a PID control module is detected. Release number 4 and later.
Security
Set this to On to load only the security settings.