2-81
IM 04L51B01-01EN
Operating the GX/GP
2
2.8 Performing File Operations
This section explains file operations.
2.8.1 Initializing the Internal Memory
You can initialize the settings. When you initialize the settings, the internal memory data is
also initialized.
For the procedure, see page 1-249 in section 1.29.1, “Initializing the Settings and the
Internal Memory”.
2.8.2 LoadingandDisplayingtheMeasuredData(Displaydataandeventdata)
from the Storage Medium
This section explains how to load display data or event data from the external storage
medium and display it as waveforms. The loaded data is shown on the historical trend
display.
For details on operating the historical trend display, see page 2-31 in section 2.2.7, “Displaying
Previously Measured Data (Historical trend display)”.
Procedure
1
Press MENU.
The menu screen appears.
2
Tap the Browse tab and then Save load.
The menu screen appears.
3
On the menu, Tap Load display data or Load event data.
4
Tap Media kind and then SD or USB.
5
Tap Select file, select the file, and tap OK.
6
Tap Execute.
The le is loaded and shown in the historical display.
Operation complete
Note
• The extensions to display data files and event data files vary depending on whether the
advanced security function (/AS option) is installed and whether the function is enabled or
disabled.
Advanced Security Function Display Data File EventDataFiles
Not available or disabled .GDS .GEV
Enabled .GSD .GSE
• You can sort the files by file name, directory name, or update date/time.
• A data file recorded in Dual interval mode contains only a single type of measurement group.
Data of measurement groups that are not recorded is not displayed.