EasyManuals Logo
Home>YOKOGAWA>Recording Equipment>GX20

YOKOGAWA GX20 User Manual

YOKOGAWA GX20
707 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
1-33
IM 04L51B01-01EN
Configuring the GX/GP and Viewing the Settings
1
2
3
4
5
App
Index
RJC
1, 3
(Referencejunctioncompensation)
Setup Item Selectable Range or Options Default Value
Mode Internal/External Internal
Temperature
2
–20.0 to 80.0°C, 253.1 to 353.2K (KpvsAu7Fe) (when the
measurement range is in Fahrenheit: –4.0 to 176.0, 253.1
to 353.2K)
0.0°C,
273.2K (KpvsAu7Fe)
1 Appears when the range type is set to TC.
2 Appears when the mode is set to External.
3 Does not appear for AI (mA) channels.
Mode
Set the reference junction compensation method of the thermocouple.
Options Description
Internal Uses the reference junction compensation function of the GX/GP.
External Uses an external reference junction compensation function.
External reference junction
Internal reference junction
GX/GP
Thermocouple
Thermocouple
Temperature
When the RJC is set to external, set the compensation temperature.
1.2ConfiguringAIChannels(Analog(includingDI)input)channelsandAI(mA)channels

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA GX20 and is the answer not in the manual?

YOKOGAWA GX20 Specifications

General IconGeneral
BrandYOKOGAWA
ModelGX20
CategoryRecording Equipment
LanguageEnglish

Related product manuals