1-268
IM 04L51B01-01EN
CommandsApplicabletoRS-422-A/RS-485Only(Releasenumber2(version2.02)and
later)
Command Description
Esc+O Open Command (address a communication destination)
EXC+C Close Command (close the addressed state of a device)
Data Output Request Commands
Command Parameter Description
TS 0: Measured data Selects the talker output data.
1: Setting parameters
2: Unit information
5: System configuration information
9: Setup mode setting data output
FM 0: Measured data (ASCII) Selects the output format of
measured/math data.
1: Math data (binary)
2: Math data (ASCII)
3: Math channel (binary)
LF — Sets the output channel for the
setting data output, unit, and
decimal place information.
CF — Sets the system configuration data
format.
VF — Relay status output request
BO — Sets the byte output order.
Esc+S
1
— Status byte output command
1 Release number 2 (version 2.02) and later
Responses to Output Request Commands
AO module channels are not output in response to the following command.
Command Setting
FM Selects the output format of measured/math data.
EF Measured data and math data (binary format)
LF Sets the output channel for the unit and decimal place information.
EL Unit and decimal point information of a specified channel (ASCII format)
Module Names Output by the CF Command
GX/GP Module Module Name
AI module INPUT
DI module DI
DO module RELAY
Expansion module Module not connected
Module error XXXXXX
DI/DO module REMOTE
Pulse input module PULSE
AO module OUTPUT
PID control module PID
1.31UsingtheDARWINCompatibleCommunicationFunction(Releasenumber2andlater)