<7. Startup>
7-4
IM 11M12G01-02EN
7.6 Conrmation of Detector Type Setting
(1) Press the [Sensor menu] key.
(2) Select [Setting] > [Setting].
(4) To change detector, press [Selection of detector]
Figure 7.4 Selection of detector
CAUTION
sensor/detector and the converter. Then change detector settings appropriately.
7.7 Selecting moisture base
Only for Oxygen Analyzer.
Combustion gas contains water vapor generated by hydrogen combustion in the fuel. Therefore,
if this water vapor is removed, the oxygen concentration at this time shows a higher value than
when water vapor is contained. Here, you can specify whether to use the value in wet gas as the
measured oxygen concentration value, or arithmetic and use it as the value in dry gas.
If you select "Choice of moisture base" in screen of the 7.6, the window "Wet", "Dry" to select.
The factory default is "Dry".
1sh Edition : Mar. 25, 2021-00