EasyManuals Logo
Home>YOKOGAWA>Measuring Instruments>ZR22G

YOKOGAWA ZR22G User Manual

YOKOGAWA ZR22G
241 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #205 background imageLoading...
Page #205 background image
<11. Inspection and Maintenance>
11-2
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
(4) Clean the U-shaped pipe. The pipe can be rinsed with water. However, it should be dried out
thoroughly before reassembly.
(5) Restore all components you removed for cleaning. Follow Section 11.1.3 to restore all
components in their original positions. Be sure to replace the O-ring(s) with new ones.
Calibration gas tube
Exploded view of components
Rod
(with outside diameter
of 2 to 2.5 mm)
Figure 11.1 Cleaning the Calibration Gas Tube
11.1.3 Replacing the Sensor Assembly
The performance of the sensor (cell) deteriorates as its surface becomes soiled during operation.
Therefore, you have to replace the sensor when its life expectancy expires, for example, when
it can no longer satisfy a zero gas ratio of 100 ± 30% or a span gas ratio of 0 ± 18%. In addition,
the sensor assembly is to be replaced if it becomes damaged and can no longer operate during
measurement.
If the sensor becomes no longer operable (for example, due to breakage), investigate the cause
and remedy the problem as much as possible to prevent recurrence.
CAUTION
• If the sensor assembly is to be replaced, allow enough time for the detector to cool down
from its high temperature. Otherwise, you may get burned.
• If the cell assembly is to be replaced, be sure to replace the metal O-ring and the contact
together.
Additionally, even in a case where the cell is not replaced, if the contact becomes deformed
and cannot make complete contact with the cell, replace the contact.
• If there is any corroded or discolored area in the metal O-ring groove in which the contact
is embedded, sand the groove with sandpaper or use a metal brush, and then sand further
with a higher grade of sandpaper (no. 1500 or so), or use an appropriate metal brush
to eliminate any sharp protrusions on the groove. The contact’s resistance should be
minimized.
• Use sensor assemblies manufactured in or after Sept. 2000: the serial number on the side
of the sensor assembly should be 0J000 or later (for example: 0K123, 1AA01 etc)

Table of Contents

Other manuals for YOKOGAWA ZR22G

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ZR22G and is the answer not in the manual?

YOKOGAWA ZR22G Specifications

General IconGeneral
BrandYOKOGAWA
ModelZR22G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals