<2.Specications>
2-1
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
2. Specications

ZR22G General use separate type detector (See Section 2.2.1)
ZO21R Probe protector (See Section 2.2.2)
ZH21B Dust protector (See Section 2.2.3)
ZR22G (0.15 m) High temperature separate type detector (See Section 2.3.1)
ZO21P Adapter for High temperature probe (See Section 2.3.2)
ZR802G Zirconia Oxygen/Humidity Analyzer, Converter (See Section 2.4)
ZA8F Flow setting unit (See Section 2.5.1)
ZR40H Automatic calibration unit (See Section 2.5.2)
ZO21S Standard gas unit (See Section 2.6)
2.1 General Specications
n Standard Specications (Oxygen Analyzer)
Measurement Object: Oxygen concentration in combustion exhaust gas and mixed gas (excluding

Measurement System: Zirconia system
Measurement Range: 0.01 to 100 vol% O
2

Setting Range: Any setting in the range of 0 to 5 through 0 to 100 vol% O
2
(in 1 vol% O
2
), or partial
range
Display Range: 0 to 100 vol% O
2
Warming-up Time: Approx. 20 min.

range from 0 to 5 vol% O
2
or more and less than 0 to 25 vol% O
2

from 0 to 25 vol% O
2
or more and up to 0 to 100 vol% O
2
range
Linearity: (Excluding standard gas tolerance and the case where the reference gas is by natural
convection) (Use oxygen of known concentration (with in the measuring range) as the zero
and span calibration gases.)

2
range and sample gas
pressure within ± 4.9 kPa

2
range and sample gas
pressure within ± 0.49 kPa

2
range and sample gas
pressure within ± 0.49 kPa
 
convection.)
Both zero and span ± 2% F.S. /month
Response Time: Response of 90% within 5 seconds.
(Measured after gas is introduced from calibration gas inlet and analog output starts
changing.)
Safety, EMC, and RoHS conformity standards of ZR22G, ZR802G
Installation altitude: 2000 m or less


Measurement category: O (other)
 


strength. Degree 2 is the normal indoor environment.

Table of Contents

Other manuals for YOKOGAWA ZR22G

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ZR22G and is the answer not in the manual?

YOKOGAWA ZR22G Specifications

General IconGeneral
BrandYOKOGAWA
ModelZR22G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals