EasyManuals Logo
Home>YOKOGAWA>Measuring Instruments>ZR22G

YOKOGAWA ZR22G User Manual

YOKOGAWA ZR22G
241 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #76 background imageLoading...
Page #76 background image
<4. Piping>
4-3
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
4.1.2 Connection to the Calibration Gas Inlet
When carrying out calibration, connect the piping (6(O.D) ~4(I.D.) mm tube) from the standard

by YOKOGAWA) on a nipple (found on the local market) as illustrated in Figure 4.2, and mount a
joint (also found on the local market) at the stop valve tip. (The stop valve may be mounted on the
detector prior to shipping the detector.)
Note: Mount the stop valve close to the detector.
F4-2E.ai
NippleStop valve
Tube connection joint
Figure 4.2 Connection to the calibration gas inlet
4.1.3 Connection to the Reference Gas Inlet
 
convection for the reference gas (models ZR22G----C). Leave the plug as it is . If the
air around the detector is polluted and the necessary oxygen concentration (21 vol%O
2
)
cannot be obtained, prepare piping the same as which described in Section 4.2, System 2.
 
described in Section 4.2, System 2 (models ZR22G----E or P).
4.1.4 Piping to the High Temperature Probe Adapter
The sample gas should be at a temperature below 700°C before reaching the detector
sensor. If the gas is under negative pressure, it should be fed to the detector by suction.
For usage of the probe adapter when using high temperature detector, refer to Section “3.2.1
Usage of the High Temperature Probe Adapter (ZO21P-H)”.
If the sample gas is under negative pressure, connect the ejector assembly (E7046EC/
E7046EN) as illustrated in Figure 4.3. Mount the pressure gauge as close as possible to
the ejector assembly. However, if the ambient temperature is too high, mount the gauge in a
location with a temperature below 40°C.
Ejector assembly for
high temperature use
Adapter for
high temperature probe
Detector
F4-3E.ai
Figure 4.3 Mounting the ejector assembly

kPa, the sample gas temperature may not be below 700°C at the detector.

Table of Contents

Other manuals for YOKOGAWA ZR22G

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ZR22G and is the answer not in the manual?

YOKOGAWA ZR22G Specifications

General IconGeneral
BrandYOKOGAWA
ModelZR22G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals