EasyManuals Logo
Home>YOKOGAWA>Measuring Instruments>ZR22G

YOKOGAWA ZR22G User Manual

YOKOGAWA ZR22G
241 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #155 background imageLoading...
Page #155 background image
<9. Calibration>
9-2
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
0.1
0.5 1 5 10 21.0
50 100
120
100
80
60
40
20
0
-20
-40
0.51 vol% O2,81.92mV(Zero origin of calibration)
21.0 vol% O2, 0mV
(Span origin of calibration)
Oxygen concentration (vol % O
2)
Cell
voltage
(mV)
F9.1E.EPS
Figure 9.1 Oxygen concentration in a Sample Gas vs Cell Voltage (21 vol%O
2
Equivalent)
The measurement principles of a zirconia oxygen analyzer have been described above.
However, the relationship between oxygen concentration and the electromotive force of a cell is
only theoretical. Usually, in practice, a sensor shows a slight deviation from the theoretical value.

is conducted so that a calibration curve is obtained, which corrects the deviation from the
theoretical cell electromotive force.
9.1.2 Measurement Principle of Zirconia Humidity Analyzer
A solid electrolyte such as zirconia allows the conduction of oxygen ions at high temperatures.
Therefore, when a zirconia-plated element with platinum electrodes on both sides is heated up
󰀨
from a high partial-oxygen pressure to a low partial-oxygen pressure, causing a voltage. When
a sample gas introduced into the zirconia-plated element with the measurement electrode, and



where, R = Gas constant
T = Absolute temperature
n: 4
F = Faraday’s constant
y = O2 vol % on the zirconia element measurement electrode
a = O2 vol % to 21.0 vol % O2 on the zirconia element reference electrode
The humidity analyzer uses a sample gas composed of water vapor and air.
(A) For the vol % H2O measurement
x:Assuming that H2O vol % in a mixed gas is measured:


Table of Contents

Other manuals for YOKOGAWA ZR22G

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA ZR22G and is the answer not in the manual?

YOKOGAWA ZR22G Specifications

General IconGeneral
BrandYOKOGAWA
ModelZR22G
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals