<10. Other Functions>
10-15
IM 11M12G01-02EN 1sh Edition : Mar. 25, 2021-00
10.6.2 Setup of Hold time, interval, start date, and start time
The stabilization time is turned setup because the measured value changes temporarily
immediately after the simple cell resistance measurement. When “Hold time” is selected, the
numeric entry screen is turned Display, so enter the hold time. Setup can range from 00 minutes
degradation occurs, the meter returns to the normal measurement in 1 minute.
When using Auto Simple Cell Resistance Measure, “Interval”, “First Start Date” and “First Start
time” must be turned setup. The term “Interval” refers to the period at which automatic simple
cell resistance measurement is performed. Select “Interval” and enter it from the numeric entry
screen. Setup is possible for 000 days 00 hours to 255 days 23 hours. For Start Date and Start
and the start time.
Note
• The degradation tendency of the cell can be checked with the simple cell resistance trend
by performing automatic simple cell resistance measurement. Setup of the automatic
simple cell resistance measurement is recommended when the measurement downtime by
executing the simple cell resistance measurement is not an issue.
resistance measurement. If the hold time is included, the measurement stops for more than
1 minute.
• Simple cell resistance measurement is not executed during calibration. When the start time
of auto-simple resistance measurement comes during calibration, the calibration ends and
the simple resistance measurement will be executed after the instrument returns to the
measurement status.
• When the start time of auto-simple cell resistance measurement comes during
maintenance, semi-auto simple cell resistance measurement, the auto simple cell
resistance measurement of this time will not be executed.
• When “000 days 00 hours” is set to setup for the interval, simple cell resistance
• If the previous date is turned setup on the starting date, the auto simple cell resistance
measurement is not executed.
10.6.3 Default
When the analyzer is delivered, or if data are initialized, the default set values are as shown in
Table 10.7.
Figure 10.7 Default value of the simple cell resistance setting
Item Default
Mode Semi-automatic
Hold time [mm:ss] 01:00
Interval [day] 1
Interval [hour] 0
First start date [YY/MM/DD] 00/01/01
First start time [hh:mm] 00:00
Simple cell resistance alarm value