EasyManuals Logo
Home>Hioki>Voice Recorder>MEMORY HiCORDER MR8880

Hioki MEMORY HiCORDER MR8880 User Manual

Hioki MEMORY HiCORDER MR8880
300 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #227 background imageLoading...
Page #227 background image
12.1 Connecting to the External Control Terminals
217
Chapter 12 External Control
12
The external control terminals on the Memory HiCorder support trigger signal input
and output.
External Control
Chapter 12
1 START/IN1
Input signals for controlling the operation of the
instrument. (Start, stop, abort, print)
Input
2 STOP/IN2
3 GND
(common with instrument ground)
4 ALM/OUT
Output signals for instrument status.
(Printing, saving, instrument alarm)
Output
5 TRIG.OUT
Output signal when trigger applied.
Output
6 EXT.TRIG
(External trigger input)
Input external signals.
Input
External control
terminals
(Right side)
12.1 Connecting to the External Control
Terminals
To avoid electrical hazards and damage to the instrument, do not
apply voltage exceeding the rated maximum to the external con
-
trol terminals.
I/O terminals Maximum input voltage
Input EXT.TRIG
0 to 10 V DC
START/IN1
STOP/IN2
Output ALM/OUT
0 to 50 V DC
TRIG.OUT

Table of Contents

Other manuals for Hioki MEMORY HiCORDER MR8880

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hioki MEMORY HiCORDER MR8880 and is the answer not in the manual?

Hioki MEMORY HiCORDER MR8880 Specifications

General IconGeneral
BrandHioki
ModelMEMORY HiCORDER MR8880
CategoryVoice Recorder
LanguageEnglish

Related product manuals