Appendix 1 Error Messages and Remedial Actions
A1
Appendix
L
Error messages consist of either "Error" or "Warning" displays.
A screen message appears whenever an error occurs. In either case, take the reme-
dial action indicated.
To clear an error message, press the ENTER key.
Appendix
Appendix 1 Error Messages and Remedial
Actions
Error Messages
Message Remedial Action
001
The internal buffer memory has been
cleared.
(Cause: Possible buffering time ex-
ceeded)
The internal buffer memory can only be backed
up more than five minutes after turning on the
power. Do not turn off the power for five minutes
after it is turned on. Once the back-up function is
activated, the internal buffer memory can be
backed up for approximately 30 minutes even af
-
ter the power is turned off.
004
Battery low. Cannot access to the
waveform backup and media.
Low battery charge: recharge or replace with a
fresh battery pack.
005
FPGA error
Incompatible FPGA firmware version. Try updat-
ing the firmware, and if the error persists, submit
the Memory HiCorder for repair.
007
Failed to update ROM.
Try updating the firmware, and if the error per-
sists, submit the Memory HiCorder for repair.
008
FPGA PLL error. Turn off and on the
power.
The instrument has malfunctioned. Turn the pow-
er off and then back on. If the error message ap-
pears after turning the power off and then back
on, the instrument needs to be repaired.
010
The amp adjustment value is abnor-
mal. This unit needs to be repaired.
There is a problem with the input section. If the er-
ror message appears after turning the power off
and then back on, the instrument needs to be re-
paired.
012
Internal memory error.
The instrument has malfunctioned. Turn the pow-
er off and then back on. If the error message ap-
pears after turning the power off and then back
on, the instrument needs to be repaired.