V Series Operating Instructions 16 - 7
ST ST Dialog
16.2.5 Reviewing ST Points
1. Select the ST tile.
2. Select the Review tab.
FIGURE 16-4 Example ST Dialog (Review Tab)
3. Select Accept or Cancel.
•The Accept button saves the settings and closes the dialog.
•The Cancel button closes the dialog without saving the settings
16.2.6 ST Relearning
NOTE: Selecting the Relearn ST button initiates the learning process for ST
analysis.
The V 12/V 21 initiates the learning process for ST or Arrhythmia analysis after one of the following:
•Unit power-up
• Return to normal monitoring from the Standby mode
• After enabling ST or Arrhythmia analysis
• The lead has been changed in ECG 1 waveform (3 lead only)
•After selecting the Relearn button in the ST, ECG, or Arrhythmia tab.
Selecting the Relearn button is recommended after one or more of the following:
• ECG electrodes have been repositioned
• Eight hours have passed since the last relearn
• After significant changes to the patient QRS complex