Continuous Cardiac Output (CCO) Performance Product Specifications
29 - 46 V Series Operating Instructions
29.11 Continuous Cardiac Output (CCO)
Performance
The VDI module has the capability to interface with Edwards Lifesciences Vigilance, Edwards
Lifesciences Vigilance 2, Edwards Lifesciences Vigileo, and Edwards Lifesciences EV1000 in Adult and
Pediatric Mode. For the performance specifications of the Edwards monitors, refer to its operator’s
manual or contact Edwards Lifesciences Corporation for assistance.
1. CCO/CCO STAT Alarms:
The VDI module is equipped with High and Low CCO/CCO STAT alarms. Alarms can be triggered
when measured rate is equal to or greater/smaller than the alarm setting.
High Alarm Limits:
Low Alarm Limits:
2. CCI/CCI STAT Alarms:
The VDI module is equipped with High and Low CCI/CCI STAT alarms. Alarms can be triggered when
measured rate is equal to or greater/smaller than the alarm setting.
High Alarm Limits:
Low Alarm Limits:
3. SVR Alarms
The VDI module is equipped with High and Low SVR alarms. Alarms can be triggered when measured
rate is equal to or greater/smaller than the alarm setting.
High Alarm Limits:
Low Alarm Limits:
Adult: Off, 2.0 to 20.0 in steps of 0.1, default Off
Pediatric: Off, 2.0 to 20.0 in steps of 0.1, default Off
Adult: Off, 1.0 to 19.0 in steps of 0.1, default Off
Pediatric: Off, 1.0 to 19.0 in steps of 0.1, default Off
Adult: Off, 2.0 to 20.0 in steps of 0.1, default Off
Pediatric: Off, 2.0 to 20.0 in steps of 0.1, default Off
Adult: Off, 1.0 to 19.0 in steps of 0.1, default Off
Pediatric: Off, 1.0 to 19.0 in steps of 0.1, default Off
Adult: Off, 500 to 5000 in steps of 1, default Off
Pediatric: Off, 500 to 5000 in steps of 1, default Off
Adult: Off, 0 to 2000 in steps of 1, default Off
Pediatric: Off, 0 to 2000 in steps of 1, default Off

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals