Product Specifications Power Specifications
29.17 Power Specifications
Power Supply Type
• AC power, battery
• Whole-system consumption: maximum of 215 W.
AC Mains Power Source
• AC power: 100~240 Vac, +/-10%
• Current: 2.4 - 1.0 Amps
• AC input power frequency: 60 Hz/50 Hz
Internal User-Removable Battery
• Li-Ion battery (0146-00-0091-01, Mindray)
• Number of batteries: 3 (maximum)
Li-Ion Battery Index
• Voltage: 11.1VDC
• Rated Battery capacity: 4.8 Ah (one battery)
• Battery run time: >240min (three batteries)
Operating conditions: 25ºC for new fully-charged battery. Equipment configuration: 5-lead ECG,
Impedance Respiration, Masimo SpO
2
, Temperature, IBP1, IBP2, and NIBP in 15-min Interval mode.
Battery charge time:
• 3.5 hours for 90% charge and 5.5 hours for full charge (equipment powered off)
• 5.5 hours for 90% charge and 8.5 hours for full charge (equipment powered on)
29.17.1 Recorder Performance Specifications
The recorder can print physiological waveforms.
Recorder paper: 50mm wide spooled thermal strip chart paper, minimum 200 ft long,
max diameter 50 mm when rolled.
Resolution: 600 dots per 25 mm at a print speed of 25mm/s, 400 dots at all other
print speeds
Maximum traces: 3 traces on thermal strip chart paper
Printing speed: 5 printing speeds: 3.125, 6.25, 12.5, 25, and 50mm/sec, each within
±
5%
Paper installation: Auto feed
Sensitivity:
± 5% on each trace
Performance: IEC60601-2-27:2011, and IEC60601-2-25:2011