V Series Operating Instructions 3 - 17
Display Setup Display Options Dialog
3.2.15 Privacy Mode
The Privacy Mode hides the patient data and suppresses alarm notification at the monitor while still
monitoring the patient. This feature is only available when a patient is connected to a central
monitoring system. The central station displays the patient data and provides alarm notification while
the patient monitor is in this mode. If communication with the central station is lost or a module is
added or removed, the monitor automatically exits the Privacy Mode. While in this mode, the display
alarm light or LED, located at the top of the monitor, may be enabled or disabled.
NOTE: Refer to "Display Options Dialog" on page 3-2 for information.
NOTE: This button is only available when the patient is being monitored at a central
monitoring station.
To enter Privacy Mode:
1. Select the Setup tab from the Navigation Area.
2. Select the Display button.
3. Check or uncheck the Disable Alarm Light check box in the Display Options dialog.
NOTE: Selecting the Disable Alarm Light check box disables the alarm lights/
indicators. Refer to "V 12/V 21 Physical Views" on page 1-4 for
additional information.
4. Select the Privacy Mode button.
The display clears and the screen reads privacy mode.
3.2.16 Screen Lock Mode
The Screen Lock mode disables the touch screen for 10 seconds. Enter this mode if the display is
being cleaned. While in this mode, a timer countdown is shown indicating how much time is
remaining.
To enter Screen Lock mode:
1. Select the Setup tab from the Navigation Area.
2. Select the Display button
3. Select the Screen Lock button.
The display indicates that the screen is locked and counts down the time remaining until it exits
the mode and returns to the Main Screen. The timeout is 10 seconds.
3.2.17 Standby Mode
Refer to "Placing a Patient in Standby" on page 6-7 for additional information.
3.2.18 Discharge Mode
Refer to "Discharging a Patient" on page 6-6 for additional information.
3.2.19 View All ECG Mode
The View All ECG mode displays all the ECG waveforms available based on the lead set currently in
use. Enter this mode when using a 5 lead or 12 lead.
To enter the View All ECG Mode:
1. Select the Setup tab from the Navigation Area.
2. Select the Display button, then select the View All ECG button.
The Display Options dialog automatically closes and the first six (6) ECG waveforms populate the
ECG waveforms tiles.

Table of Contents

Other manuals for Mindray V12

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray V12 and is the answer not in the manual?

Mindray V12 Specifications

General IconGeneral
BrandMindray
ModelV12
CategoryMedical Equipment
LanguageEnglish

Related product manuals